Recall Z-0149-2010 ยท FPP Stretcher, Hand-Carried

Stryker Medical Div. of Stryker Corporation

Stryker Medical Stair-PRO Stair Chair with wheels and a track system, patient transport device, Model 6252.

DateNov 9, 2009 (initiated Sep 2, 2009)
StatusTerminated
ReasonThe front of the handle may become detached from the chair, which could result in a patient fall or an operator injury.
Root causeDevice Design
ActionConsignees were sent a letter titled URGENT MEDICAL DEVICE CORRECTION dated 9/9/09, advising them of the problem and to remove the devices from service until they have been repaired. U.S. customers were notified that Stryker will complete the correction while international customers were sent a kit with instructions for completion of the correction. International consignees were told that Stryker will complete the correction if the consignee does not feel comfortable in doing so. Questions or comments regarding this issue should be addressed to Sarah Marie Puterbaugh of Stryker Medical at (
Quantity165
DistributionWorldwide distribution - Colorado, Ohio, Wisconsin, and Germany, Italy, Netherlands, Poland and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.