Recall Z-1630-2010 ยท GKZ Counter, Differential Cell

Abbott Laboratories

CELL-DYN 1800, list numbers 07H77-01, 07H77-03, manufactured by Abbott Hematology, Santa Clara, CA

DateMay 20, 2010 (initiated Jun 8, 2009)
StatusTerminated
ReasonUse of Syringe LN 04H36-01 with the Die-cast syringe drive may result in leakage, leading to delay in reporting patient results.
Root causeNonconforming Material/Component
ActionA Product Correction Letter and Customer Rely Form were sent to all currently active Cell -Dyn Sapphire customer. New customers are being notified with a Product Information Letter. For questions regarding this information, U,S, customers should call Customer Support at 1-877-4ABBOTT. Customers outside the U.S., please contact your local hematology customer support representative.
Quantity1805 units
DistributionWorldwide distribution: United States, Canada, Mexico, Anguilla, Australia, Argentina, Belize, Bermuda, Brazil, China, Chile, Columbia, Costa Rica, Ecuador, El Salvador, Germany, Guatemala, Haiti, Honduras, Hong Kong, Japan, Paraguay, Peru, Puerto Rico, Singapore, South Korea, Uruguay, and Venezuel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030513CELL-DYN 1800 SYSTEM
Abbott Laboratories, May 12, 2003