ClearanceBrief
Recall Z-2149-2011 ยท

Siemens Medical Solutions USA, Inc.

4Z1c Matrix Array Transducer, model number 10033682, used with ACUSON SC2000 ultrasound systems, model number 10433816, manufactured by Siemens Medical Solutions, Ultrasound Business Unit, Mountain View, CA. The 4Z1c transducer is a special purpose ultrasound transducer designed to help clinicians perform adult echocardiography and pre-operative and intra-operative surgical assessments. Secondary

DateMay 5, 2011 (initiated Jul 10, 2009)
StatusTerminated
ReasonBurn risk-- due to coolant leaks, patients may suffer burns or blisters due to increased temperature.
Root causeEquipment maintenance
ActionA customer safety advisory notification was sent out to customers on 7/10/09, and a mandatory software update will also be conducted.
Quantity42 units
DistributionNationwide. Foreign: Japan, Italy, South Korea, Singapore, Yemen, Germany, Australia, and Greece.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072365SEQUOIA PLUS DIAGNOSTIC ULTRASOUND SYSTEM
Siemens Medical Solutions USA, Inc., Ultrasound DI, Oct 17, 2007