Recall Z-2065-2011 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

Acuson X300 systems at software versions 1.00.00 to 1.0.10 and 2.0.00 to 2.0.02, model numbers 10037409, 10132987, and 10038837, manufactured by Siemens Medical Solutions ultrasound business unit, Mountain View, CA Medical device is an ultrasound imaging system for general radiology, fetal, abdominal, intraoperative, pediatric, small parts, neonatal/adult cephalic, cardiac, transesophageal, pelvi

DateApr 25, 2011 (initiated Jul 30, 2009)
StatusTerminated
ReasonSoftware Error with potential for misinterpretation of data-- The previous patient's last capture clip image can appear in the next patient's study.
Root causeSoftware design
ActionThe firm, Siemens, sent a "CUSTOMER SAFETY ADVISORY NOTIFICATION" letter to its customers on July 30, 2009. The letter described the product, problem and actions to be taken. The customers were instructed they need to update any of the patient registration settings to reset the cine buffer so that no data is transferred between studies, if the system software is not being updated. Siemens informed the customers that a mandatory update software is being installed on their system free of charge. If you have any questions, contact (650) 969-9112.
Quantity2312 units
DistributionWorldwide distribution: USA (nationwide) and countries including: Albania, algeria, Argentina, Australia, Austria, Azerbaijan, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile,China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072676ACUSON X300 ULTRASOUND SYSTEM
Siemens Medical Solutions USA, Inc., Dec 18, 2007