Recall Z-1970-2011 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

LANTIS oncology Information System, an accessory to the Linear Accelerator, as part of the COHERENCE RT Therapist, but also cleared under the ONCOR Avante-Garde and Expression, as well as the syngo RT Therapist system, as part of the ARTISTE And PRIMUS system. Manufactured by IMPAC Medical Inc., Sunnyvale, CA, device distributed by Siemens Medical Solutions Oncology Care System, Concord, CA Elec

DateApr 13, 2011 (initiated Aug 21, 2009)
StatusTerminated
ReasonWhen editing the note type of a previously entered note, then reverting to the original type, data may be lost.
Root causeSoftware design
ActionSiemens sent an Urgent Medical Device Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to configure their system to lock all new notes automatically. To automatically lock new notes, open eChart Navigator, right mouse click in the notes section, select "configuration" and select "Automatically Lock New Notes." If the customer chooses not to configure their system to automatically lock new notes, follow at least one of the suggestions listed below: If the customer must edit a note
Quantity30 affected units
DistributionWorldwide Distribution - USA including DE, WI, TX, TN, NJ, PA, NE, MA, and AL and the countries of South Africa, Australia, Ireland, Germany, Brazil, South Korea, India, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031764ONCOR AVANT-GARDE WITH COHERENCE WORKSPACES
Siemens Medical Solutions USA, Inc., Sep 5, 2003
K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006
K072485ARTISTE MV SA
Siemens Medical Solutions USA, Inc., Dec 27, 2007
K993425PRIMUS
Siemens Medical Solutions USA, Inc., Nov 26, 1999