Recall Z-0090-2010 ยท KDQ Bottle, Collection, Vacuum
Ohio Medical Corporation
Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0302-600 (6700-209-500); the bottles are sold alone and with various cap & float assemblies: 0321-7986-800 - Bottle with Straight Wall Bracket, Cap & Float; 6700-0018-800 - Bottle with V Wall Bracket, Cap & Float; 6700-0370-800 - Bottle with Floor Bracket, Cap & Float. The collection bottles ar
| Date | Oct 29, 2009 (initiated Sep 21, 2009) |
| Status | Terminated |
| Reason | The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate. |
| Root cause | Process control |
| Action | Ohio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270. |
| Quantity | 823 bottles |
| Distribution | Worldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, Philippines, Qatar, Saudi Arabia, Singapore, Thailand, Turkey, United Kingdom and Venezuela. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.