Recall Z-0089-2010 ยท KDQ Bottle, Collection, Vacuum

Ohio Medical Corporation

Vacuum Collection Bottle, 1/2 gallon (1300 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0301-600 (6700-0206-500); the bottles are sold alone and with various cap & float assemblies including: 0321-7986-801 - Bottle with Straight Wall Bracket, Cap & Float; 6700-0019-800 - Bottle with Floor Bracket, Cap & Float; 6700-0024-800 - Bottle with V Wall Bracket, Cap & Float. The collec

DateOct 29, 2009 (initiated Sep 21, 2009)
StatusTerminated
ReasonThe re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate.
Root causeProcess control
ActionOhio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270.
Quantity5,892 bottles
DistributionWorldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, Philippines, Qatar, Saudi Arabia, Singapore, Thailand, Turkey, United Kingdom and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.