Recall Z-0645-2010 ยท DPS Electrocardiograph
GE Medical Systems, LLC
GE Healthcare MAC 800 RESTING ECG ANALYSIS SYSTEM. The MAC 800 is a portable EGG acquisition, analysis and recording system. The MAC 800 is intended to acquire, analyze, display and record information from adult and pediatric populations. Pediatric population is defined as patients between the ages of 0 and 15 years. The MAC 800 is intended to be used by trained operators in a hospital or medical
| Date | Jan 21, 2010 (initiated Dec 10, 2009) |
| Status | Terminated |
| Reason | GE Healthcare has become aware of an issue with MAC800 laser printouts. When more than one ECG report in the file manager is batch printed in laser print, the potential exists for an incorrect patient identification or name to be displayed on the print out. Misidentification of patient data potentially contributing to delayed or incorrect treatment is then possible. |
| Root cause | Software design |
| Action | Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated December 10, 2009. The letter was addressed to Healthcare Administrator/Risk Manager, Chief of Nursing and Director of Biomedical Engineering. The letter described the Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. GE Healthcare will provide a software upgrade to resolve the issue. A GE Healthcare representative will contact customers to schedule implementation of the correction when it is available for installation at the customer facility. Direct quest |
| Quantity | 567 |
| Distribution | Worldwide Distribution -- United States (CA, CO, CT, FL, IL,MD, NY, NC, OH, TN, TX, WA and WI), UNITED KINGDOM, UNITED ARAB EMIRATES, THAILAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVAKIA, SINGAPORE, ROMANIA, QATAR, PORTUGAL NETHERLANDS, LITHUANIA, KOREA, JAPAN, ITALY, ISRAEL, INDIA, HUNGARY, HONG KONG, G |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K090212 | MAC 800 RESTING ECG ANALYSIS SYSTEM Ge Medical Syst. Information Technologies, Feb 19, 2009 |