Recall Z-0003-2009 ยท KPR System, X-Ray, Stationary

GE Medical Systems, LLC

Revolution XR/d systems with control room PC model 5117866-2 with Software collector 5194381. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188. Intended for use in generating Tomographic images of human anatomy.

DateSep 22, 2009 (initiated Feb 14, 2008)
StatusTerminated
ReasonGE Healthcare has recently become aware of an error in the generator software on the Revolution XR/d configured with HP 8200 PC consoles that reports a mAs reading that is higher than is actually exposed to patients.
Root causeSoftware design
ActionConsignees were sent a GE Healthcare "Product Safety Notification" letter dated March 7, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology, and Radiologists. The letter listed the Safety Issue, Affected Product, Safety Instructions, and Contact Information.
Quantity83 (63 USA, 20 OUS)
DistributionWorldwide Distribution -- USA, including states of AL, AZ, CA, CO, FL, ID, IL, IN, LA, MA, MI, MO, NY, NC, OH, OR, PA, SC, SD, TN, TX, WA, and WI and countries of Canada, China, Japan, Korea, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051967REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH TOMOSYNTHESIS
GE Healthcare, Aug 9, 2005