Recall Z-0002-2009 ยท KPR System, X-Ray, Stationary

GE Medical Systems, LLC

Revolution XR/d system control room PC model 5117866-2 with Software collector 5115434

DateSep 24, 2009 (initiated Feb 14, 2008)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential risk of observing a blank image after exposure on Revolution XR/d systems equipped with HP 8200 PC user consoles that may impact patient safety. It was reported that images acquired with off-center collimation for extremities with the position selected as Digits to front or Digits to back are displayed as blank. Investigation into th
Root causeSoftware design
ActionConsignees were sent a GE Healthcare "Product Safety Notification" letter dated March 7, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology and Radiologists. The letter listed the Safety Issue, Affected Product, Safety Instructions, Corrective Action and Contact Information.
Quantity129
DistributionNationwide Distribution -- AL, AZ, CA, CO, FL, GA, HI, IL , IN, KY, LA, MD, MA, MI,MN,MS, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OK OR, PA, SC, TN, TX, VA, WA, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051967REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH TOMOSYNTHESIS
GE Healthcare, Aug 9, 2005