Recall Z-0503-2010 ยท KWP Appliance, Fixation, Spinal Interlaminal

Alphatec Spine, Inc.

Alphatec Spine Solanas Titanium Pedicle Screw Part Number 63035-14

DateDec 14, 2009 (initiated Aug 5, 2009)
StatusTerminated
ReasonAlphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40 degrees.
Root causeProcess control
ActionAlphatec Spine telephoned consignees informing them of the above mentioned Public Reason for Recall on August 5, 2009. They were asked to check their inventory to see where this product is located. If the product had not already been used Alphatec arranged to have Federal Express contact consignee for pick-up of the product(s). They were told to expect to receive the replacement(s) within 2-3 days.
Quantity158 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052201SOLANAS POSTERIOR STABILIZATION SYSTEM
Alphatec Spine, Inc., Nov 10, 2005
K071380MODIFICATION TO SOLANAS POSTERIOR STABILIZATION SYSTEM
Alphatec Spine, Inc., Aug 20, 2007