ClearanceBrief
Recall Z-0013-2010 ยท FRN Pump, Infusion

Hospira Inc

Flexiflo Quantum Enteral Pump, List number 50596, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.

DateOct 6, 2009 (initiated Aug 11, 2009)
StatusTerminated
ReasonFire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Root causeComponent design/selection
ActionFirm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics
Quantity204,999 total
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K944669FLEXIFLO QUANTUM ENTERAL PUMP
Abbott Mfg., Inc., Jul 24, 1995