Recall Z-0179-2010 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Renal Div

System 1000 family of Hemodialysis Instruments, including the System 1000, TINA, AURORA and AltraTouch; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A. The Hemodialysis and Continuous Renal Replacement Therapy devices are used for acute and chronic hemodialysis, including high flux hemodialysis. The device is intended for use by trained operators when prescribed by a physician. The devi

DateNov 12, 2009 (initiated Aug 31, 2009)
StatusTerminated
ReasonWhen the main power is lost temporarily, a software anomaly prevents the instrument from automatically changing the ultrafiltration rate (UFR) to the minimum value as described in the Operator's manual.
Root causeSoftware design
ActionBaxter sent an Urgent Device Correction letter dated August 31, 2009 to all System 1000 Series of Hemodialysis Instruments customers via first class mail on the same date, to the attention of the Hemodialysis Administrator. The letters informed the accounts that a software anomaly in software versions 4.B.0.9, 4.B.1.1 and 4.B.1.3 was discovered when the main power was lost temporarily during a simulated treatment. The anomaly prevents the instrument from automatically changing the ultrafiltration rate (UFR) to the minimum value as described in the Operator's Manual. The accounts were informed
Quantity8,258 units
DistributionWorldwide Distribution -- United States (including Puerto Rico and the Virgin Islands), Algeria, Argentina, Austria, Bahrain, Bangladesh, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guatemala, Honduras, In
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013562AURORA SYSTEM 1000 SERIES HOME SELF-CARE SINGLE PATIENT DELIVERY SYSTEM; MODELS- SYS 1000, L3 WITH N100 ARM, 4B SOFTWARE
Baxter Healthcare Corp, Dec 12, 2002
K954987ALTRATOUCH 1000 HEMODIALYSIS DELIVERY SYSTEM
Althin Medical AB an Affiliate of Baxter Intl, Aug 8, 1996