Recall Z-2076-2009 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

Medtronic Sutureless Pump Connector Revision Kit, model 8578, Method of Sterilization: Ethylene Oxide. Medtronic, Inc., Minneapolis, MN 55432-5604 USA. The 8578 revision kit catheter connects to the catheter interface with a connector pin and strain-relief sleeve. The catheter interface connects to the pump with a sutureless pump connector. The 8578 revision kit is for use with 8709 and 87

DateSep 10, 2009 (initiated Aug 27, 2009)
StatusTerminated
ReasonCurrent labeling for Medtronic Sutureless Connector (SC) catheters and revision kits (hereafter referred to as SC catheters) incorrectly states that SC catheters are compatible with Medtronic IsoMed constant-flow infusion pumps. SC catheters are compatible with Medtronic SynchroMed II and SynchroMed EL pumps. Medtronic has determined that SC catheters are not compatible with IsoMed pumps.
Root causeDevice Design
ActionA "Medical Device Correction Notification" letter was issued on August 14, 2009 to implanting and managing physicians. The letter described the nature of the issue, scope of affected product, potential severity of the issue and provided recommendations. Physicians were instructed that Sutureless Connector (SC) catheters must not be used with IsoMed pumps for future revisions and implants. A copy of the letter was also sent to the Risk Manager at each hospital associated with physicians using SC catheters. Healthcare professionals were asked to complete and return the enclosed reply card.
Quantity20,000 of all products
DistributionWorldwide Distribution -- US (including Washington D.C. and Puerto Rico), Algeria, Aruba, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Ku