Recall Z-2075-2009 ยท LKK Pump, Infusion, Implanted, Programmable
Medtronic Neuromodulation
Medtronic Intrathecal Catheter Pump Segment Revision Kit, model 8596SC, Method of Sterilization: Ethylene Oxide. Medtronic, Inc., Minneapolis, MN 55432-5604 USA. Contents: 66-cm-pump segment with attached sutureless pump connector, Spinal segment strain-relief sleeves, Pump Segment strain-relief sleeves, connector pin. This is a component of an implantable drug infusion system, which co
| Date | Sep 10, 2009 (initiated Aug 27, 2009) |
| Status | Terminated |
| Reason | Current labeling for Medtronic Sutureless Connector (SC) catheters and revision kits (hereafter referred to as SC catheters) incorrectly states that SC catheters are compatible with Medtronic IsoMed constant-flow infusion pumps. SC catheters are compatible with Medtronic SynchroMed II and SynchroMed EL pumps. Medtronic has determined that SC catheters are not compatible with IsoMed pumps. |
| Root cause | Device Design |
| Action | A "Medical Device Correction Notification" letter was issued on August 14, 2009 to implanting and managing physicians. The letter described the nature of the issue, scope of affected product, potential severity of the issue and provided recommendations. Physicians were instructed that Sutureless Connector (SC) catheters must not be used with IsoMed pumps for future revisions and implants. A copy of the letter was also sent to the Risk Manager at each hospital associated with physicians using SC catheters. Healthcare professionals were asked to complete and return the enclosed reply card. |
| Quantity | 20,000 of all products |
| Distribution | Worldwide Distribution -- US (including Washington D.C. and Puerto Rico), Algeria, Aruba, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Ku |
| FDA record | Open on fda.gov |