Recall Z-0170-2010 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Advanced Neuromodulation Systems, Inc

Eon Rechargeable Implantable Pulse Generator (IPG) System

DateNov 10, 2009 (initiated Dec 22, 2006)
StatusTerminated
ReasonProduct's external charging system on occasion was failing to communicate with the Implantable Pulse Generator preventing re-charging of the pain relief system.
Root causeDevice Design
Action12/22/2006, Marketing Update 5015 was issued to field sales representatives providing information on improving communication between the Internal Pulse Generator and its charger. May 15, 2007, Marketing Update 5022 was issued to field sales representatives to reiterate avoiding contact between the charger antenna and conductive materials and place the preferred side of the charging antenna face-down over the implant. Troubleshooting Guide was created in January 2008 for field sales representatives and technical services departments regarding isolation of the communication issue to the extern
Quantity24,750 units total
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.