Recall Z-0170-2010 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Advanced Neuromodulation Systems, Inc
Eon Rechargeable Implantable Pulse Generator (IPG) System
| Date | Nov 10, 2009 (initiated Dec 22, 2006) |
| Status | Terminated |
| Reason | Product's external charging system on occasion was failing to communicate with the Implantable Pulse Generator preventing re-charging of the pain relief system. |
| Root cause | Device Design |
| Action | 12/22/2006, Marketing Update 5015 was issued to field sales representatives providing information on improving communication between the Internal Pulse Generator and its charger. May 15, 2007, Marketing Update 5022 was issued to field sales representatives to reiterate avoiding contact between the charger antenna and conductive materials and place the preferred side of the charging antenna face-down over the implant. Troubleshooting Guide was created in January 2008 for field sales representatives and technical services departments regarding isolation of the communication issue to the extern |
| Quantity | 24,750 units total |
| Distribution | Worldwide |
| FDA record | Open on fda.gov |