Recall Z-0037-2010 ยท CAN Regulator, Pressure, Gas Cylinder
Airgas Puritan Medical
AGI-2237 Portable 2 Cylinder Manifold Cart, for connecting medical gas cylinders.
| Date | Oct 14, 2009 (initiated Aug 6, 2009) |
| Status | Terminated |
| Reason | Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection. |
| Root cause | Component design/selection |
| Action | Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly. |
| Quantity | 208 |
| Distribution | Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.