Recall Z-0083-2010 ยท HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)

The Anspach Effort, Inc.

Foot Pedal, Foot Control for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and their accessories. Catalog Numbers: EMAX2-FP and FPLUS-FP. Indicated to cut and shape bone including bones of the spine and cranium.

DateOct 28, 2009 (initiated Aug 13, 2009)
StatusTerminated
ReasonThe Anspach Effort, Inc. is recalling the Foot Control/ Foot Pedal for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and accessories. The Foot Control Cables lacks the shield connection from the cable to the connector body. The lack of shielding may result in radiated emissions which could possibly cause interference on other electrical equipment adjacent to the foot pedal
Root causeDevice Design
ActionAnspach issued an "Urgent Product Removal" Letter dated August 17, 2009 informing customers of the affected product. Consignees are asked to screen their inventory and return any affected product including a completed Product Replacement Form to the firm. For further questions, contact Anspach at 1-800-327-6887 or 1-561-627-1080.
Quantity14
DistributionWorldwide Distribution -- Canada, Indonesia, United Kingdom, Korea and United States (NC, OK, TN, CA, KS, MI and WA).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011444ANSPACH EMAX DRILL SYSTEM
The Anspach Effort, Inc., Aug 8, 2001
K080802EMAX 2 PLUS SYSTEM, MODEL EMAX 2 PLUS MOTOR, EMAX 2 CONSOLE>FIRMWARE UPGRADE, EMAX 2 FP FOOT PEDAL
The Anspach Effort, Inc., Apr 30, 2008