ClearanceBrief
Recall Z-2061-2011 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Siemens Medical Solutions USA, Inc.

Siemens ACUSON X300 Premium Edition Systems with software versions 5.0.00 and 5.0.01 (model numbers 10348531, 10348532 and 10348533), when using the V5Ms transducer (model number 08263703) and the Plus option (10427795) Product is an ultrasound imaging system intended for General Radiology,Fetal, Abdominal, Intaoperatve, Pediatric, Small parts, Neunatal/Adult Cephalic, Cardiac, Transesophageal, Pe

DateApr 25, 2011 (initiated Jul 24, 2009)
StatusTerminated
ReasonProduct Failure-- When the 64-Channel System V5M is used in the 128-Channel mode, an over current fault occurs which shuts the system down.
Root causeSoftware design
ActionSiemens sent a Customer Safety Advisory Notification letter on July 24, 2009. The letter identified the product, the problem, and the action to be taken by the customer. A software update will be conducted in the field. Until that update occurs, customers were instructed to image without the CW doppler to avoid a system shut down. Customers were instructed to contact their service support person with any questions. For questions regarding this recall call 650-694-5993.
Quantity70 units
DistributionWorldwide Distribution - USA including CA, CT, FL, MA, and TX and the countries of United Arab Emirates, Germany, Ecuador, Spain, France, Great Britian, Saudi Arabia, Sweden, Thailand, and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072676ACUSON X300 ULTRASOUND SYSTEM
Siemens Medical Solutions USA, Inc., Dec 18, 2007