Recall Z-1404-2010 ยท NVJ Applicator, Transurethral, Radio Frequency, For Stress Urinary Incontinence In Women

Novasys Medical Inc

Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA

DateApr 16, 2010 (initiated Jul 27, 2009)
StatusTerminated
ReasonInstructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, the instructions for use were revised to emphasize potential side effects. Although Novasys has not, to date, identified a device malfunction associated with the symptoms discussed above, the investigation over time shows that there is a correlation
Root causeLabeling design
ActionA Field Action notification letter was sent to all current customers (physicians) along with the revised instructions for use, and a small supply of the revised patient information brochures. The company also updated physician information brochures and patient/physician DVDs to include the information. This was shipped to physicians in August/September of 2009. Additional copies of the Patient Brochures can be ordered by contacting Novasys Customer Service at 866-784-4777. Questions should be addressed to the firm at the same telephone number.
QuantityAs of 6/12/09 there were 436 Probes at customer accounts.
DistributionUS nationwide only. (TX, FL, CA, LA, KY, MO, GA,OR WA, HI, AZ, OK AL, TN) .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042132NOVASYS TRANSURETHRAL RF SYSTEM
Novasys Medical, Inc., Dec 21, 2004