ClearanceBrief
Recall Z-2116-2011 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

ACUSON Sequoia 12.X C512/512 SS Base System, model number 10042700. Distributed by Siemens Medical Solutions USA, Inc., Business Unit Ultrasound, Mountain View, CA. The ACUSON Sequoia is a general purpose Diagnostic Ultrasound system. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in: B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler M

DateApr 29, 2011 (initiated May 15, 2009)
StatusTerminated
ReasonShock Hazard. The manufacturer failed to set the system power switch to "ON" during the line level leakage and high pot current draw testing.
Root causeRelease of Material/Component prior to receiving test results
ActionSiemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notification letter out on May 15, 2009. Service engineers were to visit each customer to conduct necessary testing. For questions regarding this recall call 650-694-5993.
Quantity10 units
DistributionWorldwide Distribution - USA including CA, IL, MD, NC, OH, and UT and the countries of Australia, China, Denmark, Spain, South Korea, Norway, Poland, Russia and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.