ClearanceBrief
Recall Z-0003-2010 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Healthcare Inc.

HeartStart MRx (M3535A/M3536A) Defibrillator/Monitor. The HeartStart MRx/MRxE is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

DateNov 9, 2009 (initiated Aug 21, 2009)
StatusTerminated
Reasonincorrect default configuration settings, insufficient battery power as the user may not be alerted that no battery is installed or that a low battery condition exists prior to disconnecting from AC or DC power
Root causeProcess change control
ActionPhilips Healthcare issued a Urgent Voluntary Medical Device Correction Notification on August 21, 2009 advising users of the problem and Customers are asked to follow the "Action to be taken by customer/user" section of the Urgent Voluntary Medical Device Correction Notification. In this section, customers are given the instructions to correct their devices. For further information or support concerning this issue, please contact Philips representative at 1-800-722-9377.
Quantity1,523 units
DistributionNationwide Canada Australia, Austria, Beharin, BRUNEI DARUSSALAM,Chian,. Finland, France, Germany,Hong Kong, India, Indonesia, Ireland, Israel,Italy, Japan, Libya, Lithuania. Mexico, Nambia, Netherlands, Palestine, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, South Africa, South Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003