Recall Z-2232-2009 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Oridion Medical 1987 Ltd.

Oridion Medical Capnostream20 capnograph/pulse oximeter monitor and monitor with printer. Part numbers CS 08651, CS 08653, CS08657, and CS08659.

DateSep 21, 2009 (initiated Jun 30, 2009)
StatusTerminated
Reason1) The monitor's ability to withstand electrostatic discharge (ESD) to exposed metal connectors on the back of the monitor which may cause the monitor's display to freeze. 2) The unit generates electromagnetic radiation above the limits set by applicable standards for medical devices
Root causeDevice Design
ActionOridian sent a letter to all its distributors June 25, 2009, instructing them to gather all the units to be exchanged for factory reconditioned units that have the hardware and software enhancements and meet all performance requirements in full.
Quantity326 units
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060065CAPNOSTREAM20 TWO PARAMETER BEDSIDE MONITOR & CAPNOSTREAM10 ONE PARAMETER BEDSIDE MONITOR
Oridion Capnography, Inc., May 4, 2006