ClearanceBrief
Recall Z-1145-2010 ยท DHC System, Test, Systemic Lupus Erythematosus

Bio-Rad Laboratories Inc

BIO-RAD 425-2020 Autoimmune EIA Anti-Cardiolipin IgM, 96 Tests...2010 02 16...Lot CL-863. Positive Control included in lot is identified as "REF 425-2025" and "LOT CPM-591". in vitro diagnostic.

DateMar 16, 2010 (initiated Jul 14, 2009)
StatusTerminated
ReasonThe firm received customer complaints of microbial contamination in Positive Controls in its Anti-Cardiolipin IgM Test Kit Lot CL-863.
Root causeNonconforming Material/Component
ActionBio-Rad initiated its field correction on July 14, 2009. The firm notified domestic consignees by phone followed by fax, and each subsidiary will be given an urgent notice, subsidiary response form, customer medical device correction notification, acknowledgement form and a listing of consignees. The subsidiaries must contact customers as instructed.
Quantity213 kits
DistributionWorldwide distribution: USA, Australia, Canada, France, Italy, Korea, Mexico, Panama, Singapore, and Trinidad.
FDA recordOpen on fda.gov