Recall Z-0011-2010 ยท FRN Pump, Infusion
Hospira Inc
AC Power Cords, shipped between 06/2005 and 05/2009. These cords were used in the following infusion pump models: Nutrimix, Q-Vue, Q2, Q2 Plus, Ox4, PCA, PCA II, PCA 3, Lifecare PCA, Lifecare 4100, Plum XL 3, Plum XL3M, Plum XL3D, Plum A+/ Plum A+3, and Gemstar Docking Station devices. AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.
| Date | Oct 6, 2009 (initiated Aug 11, 2009) |
| Status | Terminated |
| Reason | Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard. |
| Root cause | Component design/selection |
| Action | Firm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics |
| Quantity | 204,999 total |
| Distribution | Worldwide Distribution |
| FDA record | Open on fda.gov |