Recall Z-2038-2009 ยท ECX Cylinder, Compressed Gas, And Valve
Allied Healthcare Products Inc
First Responder Kit containing a 240-liter aluminum oxygen cylinder with individually packaged regulator, cannula, and bag mask resuscitator, all packaged in a shipping container labeled as: Product #L903, Desc FIRST RESPONDER KIT, Catalog #L903. The brand and firm name on the label affixed to the cylinder is Chemetron, Allied Healthcare Products Inc., St. Louis, MO.
| Date | Sep 10, 2009 (initiated May 15, 2009) |
| Status | Terminated |
| Reason | Medical valves do not meet quality specifications which may cause the valve to fail and result in uncontrolled release of oxygen when the valve is in the open position. |
| Root cause | Nonconforming Material/Component |
| Action | The recalling firm first notified their customers via a letter dated 5/15/09 issued regular mail explaining the reason for recall and requesting the cylinders with specific post valves be returned. A second letter dated 6/1/09 was issued via regular mail which provided specific shipping dates that each customer received the suspect product. Neither of these two letters requested subrecall. A third letter dated 6/17/09, which included copies of the first two letters, via certified mail. This letter informed the customer the recall was to be conducted to the user level and requested their |
| Quantity | 9 cylinders |
| Distribution | Class II Recall Worldwide Distribution including states of NY, NC, KY, PA, WI, NH, NJ, CA, CT, FL and Singapore. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.