Recall Z-1991-2009 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc.

EasyLink Informatics System included with the Dimension Vista System. The Dimension Vista Integrated System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.

DateSep 29, 2009 (initiated Jul 14, 2009)
StatusTerminated
Reasonincorrect results may be transmitted to LIS
Root causeSoftware design
ActionSiemens issued an "Urgent Device Correction" letter dated July 2009. The letter informed the user of the affected product including instructions for avoiding the problem. For further information, contact Siemens Healthcare Diagnostics Technical Solution Center at 1-800-441-9250.
Quantity359 units
DistributionWorldwide Distribution -- United States, Australia, Austria, Belgium, Canada, France, Germany, Italy, Japan, Malaysia, Netherlands, Portugal, Slovenia, Spain and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051087DIMENSION VISTA INTEGRATED SYSTEM
Dade Behring, Inc., Jul 11, 2005