Recall Z-0061-2010 ยท CAN Regulator, Pressure, Gas Cylinder

Western - a Scott Fetzer Company

Medical Gas Yoke Assembly. Airgas Portable 2-Cylinder Manifold Cart Model #AGI-2237. The device is a medical gas yoke assembly which is used to connect gas cylinder post valves to equipment used for anesthesia or respiratory therapy.

DateOct 15, 2009 (initiated Aug 10, 2009)
StatusTerminated
ReasonThe yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become compromised and allow a post valve of one gas type to become interchangeable with a yoke connection of another gas type.
Root causeComponent design/selection
ActionWestern Enterprises issued an "Urgent-Device Recall" notification dated August 20, 2009 informing Consignees of the affected product. Consignees are to contact the firm to request backup plates and installation instructions or request an RGA and shipping instructions for return of the manifold block assembly for modification. For further information, contact Western at 1-800-783-7890.
Quantity42 cart assemblies containing 84 yoke assemblies (42 Oxygen yokes and 42 Nitrous Oxide yokes)
DistributionNationwide Distribution -- AL, AR, CA, CO, FL, GA, HI, IL, IN, MA, OH, OR, PA, TX, WA and WI.