Recall Z-1953-2009 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Medtronic Inc. Cardiac Rhythm Managment

Medtronic Virtuoso DR, Model D154AWG, Dual Chamber implantable cardioverter defibrillator with atrial and ventricular therapies. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. This device is indicated to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias in patients with NYHA func

DateAug 18, 2009 (initiated Jul 27, 2009)
StatusTerminated
ReasonMedtronic has detected a specific pattern of MOSFET IC malfunctions in its Concerto, Virtuoso and EnRhythm family of devices. The probability of occurrence decreases with time and, to date over 90% of the malfunctions related to the pattern have occurred within the first twelve months after implant.
Root causeProcess control
ActionMedtronic, Inc. personnel retrieved 2 non-implanted devices remaining at 2 hospitals. A Performance Notes letter was left at each facility. The Performance Notes described the issue, actions to address issue and recommendations to physicians. The Performance Notes will also be posted on the firm's website. Direct questions about this recall to Medtronic, Inc. by calling 1-763-514-4000.
Quantity1
DistributionNationwide Distribution-- (states of IA and NJ).