Recall Z-1693-2009 ยท FMT Warmer, Infant Radiant

Draeger Medical Systems, Inc.

Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.

DateAug 26, 2009 (initiated Jul 29, 2009)
StatusTerminated
ReasonSpark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly.
Root causeProcess control
ActionDraeger Medical issued an Urgent Medical Device Recall letter via certified mail on 8/7/09. The letter explained the problem, the need to conduct an inspection of the infant warmer (instructions included), and request replacement elements.
Quantity2915 elements
DistributionWorldwide distribution: USA, Canada, Saudi Arabia, Bermuda, Virgin Islands, Mexico, Panama, Colombia, and Costa Rica.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003335RESUSCITAIRE RADIANT WARMER; RESUSCITAIRE BIRTHING ROOM WARMER; RESUSCITAIR WALL MOUNTED RADIANT WARMER
Hill-Rom Air-Shields, Nov 17, 2000