ClearanceBrief
Recall Z-1166-2010 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Medical Systems Inc

Aquilion 32 CT System; Model number TSX-101 A/I The product is an x-ray device that uses digital geometry to generate a three-dimensional image of the inside of an object from a large series of two-dimensional X-ray images take around a single axis of rotation

DateMar 19, 2010 (initiated Jan 8, 2009)
StatusTerminated
ReasonToshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large volumes of data are being reconstructed on the secondary console.
Root causeComponent design/selection
ActionCustomer notification was sent and a software upgrade will be issued to all customers affected by the field correction. If you have questions, please call (800) 521-1968.
Quantity94
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080211MODEL TSX-101A/H/I AQUILION 32/64 SP CT SYSTEM
Toshiba America Medical Systems, In.C, Mar 14, 2008