ClearanceBrief
Recall Z-2007-2009 ยท HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment

Linvatec Corp. dba ConMed Linvatec

MC5057 12ft/3.6 meters Universal Cable. Made in USA. NON STERILE Rx Only. ConMed Linvatec, 11311 Concept Blvd, Largo, FL 33773. The MC5057 Universal Cable, 12 ft, is used to connect electric handpieces to the power source. The cable is designed with a button feature that allows the user to increase or decrease the operating speed of the handpiece in use. In select handpieces, the cord button

DateOct 1, 2009 (initiated Jul 31, 2009)
StatusTerminated
ReasonSufficiently worn/damaged Universal Cable with product number MC5057, when used with various ConMed Linvatec electric Handpieces, may cause the handpiece to self activate.
Root causeComponent design/selection
ActionConMed Linvatec notified customers via letter, cautioning the users to inspect cords for signs of excessive wear or damage and urging them to discontinue use and return cables when appropriate. Direct questions to Linvatec Corporation by calling 1-800-237-0169.
Quantity12,205 units
DistributionWorldwide Distribution including United States.
FDA recordOpen on fda.gov

510(k) clearances named in this recall