Recall Z-2209-2009 ยท MNB Device, Thermal Ablation, Endometrial

Boston Scientific Corporation

Boston Scientific HydroThermAblator Endometrial Ablation System- HTA System Procedure Set with Tenaculum Stabilizer and Disposable Heater Canister (5/bx) Material/UPN/Catalog Number: M006560201 System-HTA System Procedure Set with Tenaculum Stabilizer and Disposable Heater Canister (ea) Material/UPN/Catalog Number: M006560200 Intended to ablate the endometrial lining of the uterus in premenopa

DateSep 22, 2009 (initiated Jul 31, 2009)
StatusTerminated
ReasonCracked procedure sheaths, incorrect care/use of device, and console malfunctions may result in fluid leaks and loss of a cervical seal that are responsible for burns.
Root causeProcess design
ActionBoston Scientific issued an Urgent Medical Device Recall - Immediate Action Required notification to hospitals and distributors dated July 31, 2009 via Federal Express and requested the immediate discontinued use and segregation of Recall product for return and Complete and return the Reply Verification Tracking Form. An email was sent to Physicians to provide the recall information. Boston Scientific issued a Clarification letter dated August 6, 2009 via Federal Express as a follow-up to the July 31, 2009 Recall letter for the following reason: customers may remove product from the outer bo
Quantity22,744 eaches
DistributionWorldwide Distribution -- USA, Austria, Denmark, Finland, France, Great Britain, Italy, Netherlands, and Sweden.
FDA recordOpen on fda.gov