Recall Z-1900-2009 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ArthroCare Corporation

ArthroWand Saber 30 with Integrated Cable wand, Catalog number AC4330-01, Manufactured by ArthroCare Corporation, Sunnyvale, CA

DateAug 30, 2009 (initiated Aug 5, 2008)
StatusTerminated
ReasonPotential Sterility Loss-- Due to wearing or puncture in the Tyvek lid, the product may lose sterility.
Root causePackage design/selection
ActionFirm initiated this recall on August 5, 2008. Consignees were notified via letter asking them to contact consignees and delivered via courier. Distribution discontinued on July 9, 2008.
Quantity23,443 units (includes distribution of previously returned product)
DistributionWorldwide Distribution including USA and countries of Netherlands, Slovak Republick, Ireland, Spain, Switzerland, Singapore, Russian Federatin, Greece, Japan, Saudi Arabia, India, Slovenia, Israel, South Africa, South Korea, Poland, Thailand, Jordan, Belgium, Finland, Portugal, Cyprus, Italy, UAE,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030551ARTHROCARE CONTROLLER (SYSTEM 2000 AND 8000); ARTHROCARE PATIENT CABLE; FOOT CONTROL; POWER CORD; WANDS
Arthrocare Corp., Mar 7, 2003