ClearanceBrief
Recall Z-1955-2009 ยท FMT Warmer, Infant Radiant

Draeger Medical Systems, Inc.

Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.

DateAug 25, 2009 (initiated Jul 17, 2009)
StatusTerminated
ReasonHot particles fell into the infant bassinet beneath a Radiant Infant Warmer. The resulting fire injured an infant. These devices are obsolete.
Root causeEquipment maintenance
ActionOn 7/20/09, Draeger Medical, the recalling firm, issued a press release to medical media outreach organizations and an Urgent Medical Device letter to all known customers. This information explained the incident and the need to remove and replace the units in question as soon as feasibly possible.
Quantity6489 units
DistributionWorldwide distribution: USA, Brazil, Canada, China, Dominican Republic, France, Korea, Lebanon, Mexico, Nigeria, Panama, Poland, Saudi Arabia, Taiwan, Turkey, United Kingdom, Uruguay, and Venezuela.
FDA recordOpen on fda.gov