ClearanceBrief
Recall Z-1804-2009 ยท MAF Stent, Coronary

Abbott Vascular-Cardiac Therapies dba Guidant Corporation

POWERSAIL Coronary Dilatation Catheter: 3.25x18mm (US) Part #1005524-18

DateJul 29, 2009 (initiated Jun 17, 2009)
StatusTerminated
ReasonDistal shaft of the POWERSAIL coronary Dilatation Catheters exhibits damage that may result in a leak of contrast material and functional failures.
Root causeProcess control
ActionAbbott Vascular sent an Urgent Device Recall letter (w/return form) to customers on June 17, 2009 requesting return of identified products to Abbott Vascular Territory Manager. Abbott Vascular issued a press release dated July 28, 2009.
Quantity557 total devices
DistributionWorldwide: United States, Australia, Belguim, Canada, China, Czech Republic, France, Germany, Greece, India, Italy, Japan, Jordan, Mexico, Netherlands, Norway, Poland, Puerto Rico, Spain, Sri Lanka, Switzerland, Thailand, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.