Recall Z-2050-2009 ยท BSO Catheter, Conduction, Anesthetic

Smiths Medical ASD, Inc.

Portex Continuous Epidural Tray, Reference 4037-20 An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia.

DateSep 3, 2009 (initiated Jul 15, 2009)
StatusTerminated
ReasonPortex Continuous Epidural Tray was distributed with a 20G closed end, three eyed, nylon catheter without a stylet instead of the 20 G open end Teflon catheter with stylet specified on the label.
Root causeMixed-up of materials/components
ActionSmiths Medical notified consignees by an Urgent Device Recall letter dated July 15, 2009, to locate and quarantine the affected product, stop distribution, and complete and fax back the attached Distributor Reply Letter to Smiths Medical. Credit or replacement will be arranged as requested for the return of any affected product. All distributors will be required to return any affected product in stock. All of the distributors will be required to send their customer list to Smiths Medical who will notify their customers of this recall.
Quantity190 kits (19 cases)
DistributionNationwide Distribution -- including states of New York, Texas, Ohio, Utah, Alabama, Arizona, Kentucky, & Wyoming.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062005PORTEX EPIDURAL CATHETER
Smiths Medical Asd, Inc., Oct 18, 2006