Recall Z-1912-2009 ยท GAA Needle, Aspiration And Injection, Disposable

Vital Signs Inc

Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color Zone: Green, Single Patient use, INTRAOSSEUS Module, Product # 7700GIO, Vital Signs, Totowa, NJ. The Broselow/Hinkle Pediatric Emergency System is a convenience kit that contains various modules/components used for the emergency treatment of

DateSep 8, 2009 (initiated Jun 30, 2009)
StatusTerminated
ReasonNot compatible: The luer connector on the Intraosseus needle provided does not fit securely on the luer connection of the IV Extension Set in the module.
Root causeDevice Design
ActionVital Signs, Inc. issued an "Urgent Medical Device Correction" letter dated June 30, 2009 to all direct and indirect accounts by first class mail. Consignees were informed of the affected device, asked to complete a Confirmation form for return and to return unused product for replacement. For further questions, contact Vital Signs, Inc. at 1-973-790-1330 extension 404.
Quantity23,429 kits
DistributionWorldwide Distribution -- US, Canada, United Arab Emirate, Venezuela, Guam, Korea and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.