Recall Z-1995-2009 ยท MLD Monitor, St Segment With Alarm

GE Medical Systems Information Tecnology

Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU01, L-CICU02A, L-ICU05 & L-ICUO5A software. CCCM Indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinavregional perfusion, B ispectral index (BIS), and neurophysiological status of all hospital patients.

DateSep 1, 2009 (initiated Feb 6, 2006)
StatusTerminated
ReasonGE Healthcare (Datex-Ohmeda) S/5 Critical Care Monitor and S/5 Compact Critical Care Monitor with L-CICU01, L-CICU02A, L-ICU05 & L-ICUO5A software had an issue where the Heart Rate (HR) was displaying as ---(3 dashes) when the monitor was used continuously for 49 days without a cold start. A missed alarm or missed HR value could cause a delay in recognition of a cardiac event that may result i
Root causeSoftware design
ActionGE Healthcare Technologies issued a Marketing Bulletin "Ref: MB MS HKI 06007" dated February 6, 2006 to explain a software update.
Quantity507
DistributionITALY, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRASIL, BULGARIA, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA, IRAN, ISRAEL JORDAN, LITHUANIA, ITALY, MACEDONIA, MALAYSIA, MEXICO, NETHERLAND, NICARAGUA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022740DATEX-OHMEDA S/5 COMPACT CRITICAL CARE MONITOR WITH L-CICU02 & L-CICU02A SOFTWARE
Datex-Ohmeda, Nov 14, 2002