Recall Z-1889-2009 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

NX 2008 Central Monitoring System (CMS). Product code: E2FB6 Agfa's Computed Radiography Systems with NX Workstations are intended for use in providing diagnostic quality images to aid the physician with diagnosis.

DateSep 2, 2009 (initiated Jun 22, 2009)
StatusTerminated
ReasonAfter opening and closing a study on a Central Monitoring System while a study is still open on an in-room NX system, a problem of image mix-up can occur on the in-room NX system.
Root causeSoftware design
ActionAn "Urgent Safety Notice" was sent to all potentially affected sites on/about June 22, 2009 to make them aware of the issue. A FAX Back form is requested with each letter to ensure someone at the account acknowledges the information was received and understood. If a FAX Back form is not received, the account will be contacted by telephone. A software service update to solve this issue is expected to be released by August 15, 2009. A Mandatory Service Bulletin will be deployed to correct the problem in the software. If you have questions, please contact Agfa Healthcare at 1-800-581-2432 prompt
Quantity97 units
DistributionWorldwide - IL, PA, TX, Canada, Europe, Asia Pacific and Latin America
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090672AGFA COMPUTED RADIOGRAPHY SYSTEMS WITH NX WORKSTATIONS
Agfa Healthcare N.V., May 18, 2009