Recall Z-1915-2009 ยท IQO Device, Prosthesis Alignment

Stryker Howmedica Osteonics Corp.

Stryker Orthopaedics PAL Pelvic Alignment Level, Ref: PAL 400; Howmedica Osteonics Corp, 325 Corporate Drive, Mahwah, NJ 07430. Intended to determine proper anatomical orientation, selection, alignment and position of the acetabular and femoral components.

DateAug 26, 2009 (initiated Sep 11, 2007)
StatusTerminated
ReasonThe bubble in a vial of the Pelvic Alignment Level (PAL) was wider than the level marking. Upon opening the PAL and removing the vial, it was discovered that the vial was leaking
Root causeDevice Design
ActionStryker Orthopaedics issued an "Important Product Removal" letter dated September 11, 2007 via Federal Express informing consignees of the affected devices. The firm requested that users identify and return any affected product and account for product by returning a Product Accountability Form via fax. For further information, contact Stryker Orthopaedics at 1-201-831-5825.
Quantity615
DistributionNationwide Distribution
FDA recordOpen on fda.gov