Recall Z-1810-2009 ยท IYE Accelerator, Linear, Medical

Accuray Inc

CyberKnife Treatment Delivery System, a subsystem of the CyberKnife Robotic Radiosurgery System, manufactured by Accuray Inc., Sunnyvale, CA. Indicated for treatment for planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

DateSep 14, 2009 (initiated Jun 22, 2009)
StatusTerminated
ReasonTargeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area.
Root causeSoftware design
ActionAccuray, Inc. issued an "Urgent Device Correction" notice dated June 22, 2009. The consignees were informed of the defect including methods for detecting the error. The firm is in the process of developing a software patch to address the issue. For further information, contact Accuray, Inc. Customer Support at +1-877-668-8667 (US) or +1-408-716-4700 (outside US) or customersupport@accuray.com.
Quantity61 units
DistributionWorldwide Distribution -- US including states of MA, FL, AL, TN, DE, CA, NJ, OK, CO, NC, IL, PA, MD, DC, MS, CN, NV, RI, SC, UT, VA, MI, NE, MO, TX and AL and countries of Canada, Russia, UK, Ukraine, Saudi Arabia, Switzerland, Turkey, Taiwan, India, China, Korea and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072504CYBERKNIFE ROBOTIC RADIOSURGERY SYSTEM
Accuray Incorporated, Sep 21, 2007