Recall Z-1913-2009 ยท DYB Introducer, Catheter

Edwards Lifesciences, Llc

Edwards LifeSciences, Percutaneous Sheath Introducer Set, Rx only, For single use only. Indicated for use in patients requiring access of the venous system or to facilitate catheter insertion.

DateAug 25, 2009 (initiated May 11, 2009)
StatusTerminated
ReasonEdwards LifeSciences LLC received several complaints, in which the blue connector of the introducer assembly has broken (fracture/separation) after several hours of use.
Root causeNonconforming Material/Component
ActionEdwards LifeSciences LLC initiated the worldwide recall via customer letters dated May 11, 2009 and are being sent by Federal Express stating that customers cease using affected product from the indicated lots and return all unused products. The firm requested that the customer provide written verification of stock and identify any unused product to be returned. For further information, contact Edwards LifeSciences LLC Customer Service at 1-800-424-3278.
Quantity67,357 Units
DistributionWorldwide Distribution: United States (AL, AR, CA, CO, CT, DC, FL, GA, IA, IL, IN,KS, KY, MA, MD, ME, MI, MO, MS, ND, NH, NJ, NM, NV, NY, OH, PA, RI, SC, TN, TX, VA, WV, HI and PR), Austria, Canada, Switzerland, Czech,republic, Germany, Denmark, Algeria, Estonia, Egypt, Spain, Finland, France, Great
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981909BAXTER HEMOSTASIS VALVE INTRODUCERS
Baxter Edwards, Nov 24, 1998