ClearanceBrief
Recall Z-1961-2009 ยท LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

B. Braun Interventional Systems

Celsite Access Port Model ST305P. The Celsite Access Port model ST305P is a member of the Celsite product line of implantable port and catheter systems.

DateSep 3, 2009 (initiated Jun 22, 2009)
StatusTerminated
ReasonB. Braun Interventional Systems was notified by B. Braun Medical France, the manufacturer of the Celsite Access port that they had received 11 reports of polyurethane (type P) catheter disconnections. B. Braun Medical France has decided to recall all Celsite ports supplied with the affected catheter lot. To date, B. Braun Interventional Systems (US) has not received any reports of catheter dis
Root causeNonconforming Material/Component
ActionB. Braun Interventional Systems, Inc. issued an "Urgent: Medical Device Recall" notice dated June 22, 2009. The letter identified the problem and product. The letter reminded the consignee of the general instruction for use of access ports and instructed them to contact patients who had the Celsite port system implanted. For further questions, contact B. Braun Interventional Systems, Inc. at 1-800-523-9676 extension 4818.
Quantity40 in US
DistributionNationwide Distribution -- NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061424CELSITE/ CELSITE CONCEPT, ACCESS PORTS
Aesculap, Inc., Jul 31, 2006
K952548CELSITE VENOUS SYSTEM
B.Braun Medical, Inc., Jul 21, 1995