Recall Z-2225-2009 ยท JAK System, X-Ray, Tomography, Computed

GE Medical Systems, LLC

Sytec 6000, Sytec 8000, and Lemage/ProSeed/ProSpeed families. The CT/e and HiSpeed X/i families CT Scanner Systems are indicated for head and whole body X-ray Computed Tomography applications.

DateSep 18, 2009 (initiated Dec 1, 2008)
StatusTerminated
ReasonDuring internal testing, GE Healthcare has recently determined that the collimator used on our CT/e and HiSpeed X/i families of Computed Tomography systems does not comply with the specific requirements in United States Title 21 of the Code of Federal Regulations, and IEC 60601-1-3, related to leakage radiation from the diagnostic source assembly, which includes the collimator. The excess leak
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare issued an "Important - Electronic Product Radiation Warning" to affected customers. GE Healthcare field service representative will perform a field corrective action on all affected units. All potential users of the device should be made aware of the notification and the information contained. For further information, contact GE Healthcare at 1-800-437-1171 (US) or by contacting your local GE Healthcare service manager.
Quantity643 units ( 136 US; 507 OUS)
DistributionWorldwide Distribution -- UNITED STATES (AL, AZ, AR, CA, CT, DC, FL, GA, IL, IN, KS, KY, LA, MD, MA, MI, LMN, MS, MO, NE, NJ, NM, NY, NC, OH, OK, PA, PR,SC, SD, TN, TX, VA, WA, WV and WI), ALGERIA, ANTIGUA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BENIN, BOLIVIA, BOSNIA & HERZEGOVINIA, BR