ClearanceBrief
Recall Z-1893-2009 ยท KDJ Set, Administration, For Peritoneal Dialysis, Disposable

Baxter Healthcare Corp.

Baxter Minicap Extended Life PD Transfer Set (6") with Twist Clamp; six sets per carton; Baxter Healthcare Corporation, Deerfield, IL 60015 USA; product code 5C4482 The set is used in disconnect applications and cycler applications where aseptic connections and disconnections are performed at the transfer set/cycler juncture.

DateAug 31, 2009 (initiated Jul 2, 2009)
StatusTerminated
ReasonThe peritoneal dialysis transfer set has a defective white sleeve twist clamp. The twist clamp will not open or close completely, which could cause leakage and lead to peritonitis.
Root causeNonconforming Material/Component
ActionBaxter Healthcare Corporation telephoned the accounts receiving the affected lots and then faxed or e-mailed "Urgent Product Recall" letters dated July 2, 2009, to them, to the attention of the Peritoneal Dialysis Nurse, to inform them of the defective white sleeve twist clamp, which may result in leakage and possible peritonitis. The accounts were asked to immediately discontinue the use of the affected lot numbers and examine their inventory for the affected lots. If any affected product was found, they were requested to quarantine it and contact Baxter Center for Service at 1-888-229-0001
Quantity714 sets
DistributionUS (states of Arkansas, Illinois, Mississippi, North Carolina, Tennessee and Virginia) and El Salvador.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K882498EXTENDED LIFE CAPD TRANSFER SET 5C4444
Baxter Healthcare Corp, Jul 13, 1988