Recall Z-2005-2009 ยท NBW System, Test, Blood Glucose, Over The Counter

Roche Diagnostics Corp.

Roche ACCU-CHEK Performa blood glucose test strips, Roche Diagnostics, Indianapolis, IN. Test strips for quantitative blood glucose measurements in fresh capillary, arterial, or neonatal blood, as well as in heparin (lithium or sodium) or EDTA anticoagulated venous blood.

DateSep 16, 2009 (initiated Jun 10, 2009)
StatusTerminated
ReasonThere may be a significant negative bias when testing neonate samples.
Root causeProcess control
ActionRoche Diagnostics notified consignees in the affected countries of the problem by "Urgent Field Safety Notice" dated June 10, 2009 and instructed to cease using the Accu-Chek Performa and Accu-Chek Inform II systems on neonates, except for systems using test strips with advanced chemistry. For questions and additional information, contact your local Roche Diagnostics sales affiliate or sales representative.
Quantity1,518 of total products
DistributionInternational Distribution -- Argentina, Australia, Brazil, Chili, China, Colombia, Ecuador, France, Germany, Great Britain, Hong Kong, India, Korea, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Singapore, South Africa, Taiwan, Thailand, Uruguay and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070585ACCU-CHEK PERFORMA SYSTEM
Roche Diagnostics Corp., May 25, 2007