Recall Z-1813-2009 ยท CBK Ventilator, Continuous, Facility Use

Respironics California Inc

Respironics Esprit Ventilator System, Model Number V1000 with Optional External Battery; Product Number: 1001456.

DateAug 11, 2009 (initiated Apr 22, 2009)
StatusTerminated
ReasonThe recall was initiated because through routine product monitoring Respironics discovered that under certain conditions, when operating on AC line power, a portion of the current from the External Battery charger is erroneously diverted to the ventilator. When this occurs, the External Battery charger may not transition correctly. Prolonged operation in this condition results in overcharging and
Root causeDevice Design
ActionRespironics California, Inc. issued an "Urgent - Medical Device Field Correction" notice dated April 22, 2009. Customers were informed of the affected device including corrective actions taken by the firm. A Respironics representative will contact customers to schedule service of the device. Further questions can be addressed by contacting Respironics US Customer Service at 1-800-345-6443.
Quantity508
DistributionWorldwide Distribution: United States, Hong Kong, India, Mexico, South Africa and Vietnam.
FDA recordOpen on fda.gov