Recall Z-2055-2009 ยท ITX Transducer, Ultrasonic, Diagnostic

Hitachi Medical Systems America Inc

EUB-9500/HI VISION 8500 Diagnostic Ultrasound Scanners, DICOM Option EZU-FC6, Software Versions: V11-01, V11-03, V12-03 and V12-04A.

DateSep 9, 2009 (initiated May 12, 2008)
StatusTerminated
ReasonThe firm's parent company discovered a DICOM UID (Universal Identifier) code mix-up which can occur due to a software error. This error can cause the images from two (2) different patient scans to become intermingled and consequently, labeled incorrectly.
Root causeSoftware design
ActionHMSA issued a 'DEVICE CORRECTION' notification to their customers on 5/12/2008. The notification describes the potential data transmission problem (potential to archive studies done on different patients into the same patient folder) and the conditions under which this problem can occur. The notification cautions the user/customer to verify all images stored on the PACS (Picture Archiving and Communications System) in order to ensure that they do not contain images from a different patient. The DEVICE CORRECTION letter also states that the firm will be installing updated software on all affe
Quantity6 units
DistributionThe recalled device system units were distributed to customers located in the following states within the U.S ony.: AR, CA, and TX.