Recall Z-0279-2010 ยท FPO Stretcher, Wheeled
Stryker Medical Div. of Stryker Corporation
Stryker Medical M-Series stretcher, Stryker Medical, Portage, MI; Model 1007 (SM104). Wheeled hospital stretcher.
| Date | Nov 17, 2009 (initiated Aug 10, 2009) |
| Status | Terminated |
| Reason | The brake/steer pedal may seize, resulting in the brakes not working. |
| Root cause | Nonconforming Material/Component |
| Action | Consignees were notified by letter dated August 10, 2009 that Stryker will contact them to upgrade the stretchers within 6 months. The recall was expanded in scope to include additional serial numbers by a second letter dated September 17, 2009, which was issued to additional consignees who had not been previously notified of the recall, as well as the consignees who had been issued the August 10, 2009 letter. |
| Quantity | 8825 |
| Distribution | Nationwide, Argentina, Australia, Brazil, Canada, Chile, China, France, Germany, Greece, India, Italy, Japan, Korea, Latin America, Malaysia, Mexico, Netherlands, Philippines, Poland, Portugal, Singapore, South Africa, Spain, Switzerland, Taiwan, and United Kingdom, |